Health product recall: Optima XR220amx X-Ray System

eAwazHealth

Product
Optima XR220amx X-Ray System
Issue
Medical devices – Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Issue

GE Healthcare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the column safety mechanism can be missing.

Recall start date: December 4, 2024

Additional information

Details
Original published date: 
Alert / recall type
Health product recall
Category
Health products – Medical devices – Radiology
Companies

GE Medical Systems, LLC

3000 North Grandview Blvd., Waukesha, Wisconsin, United States, 53188

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76714