Health product recall – Accu-Chek Guide

eAwazMedicine

Ottawa – The issue arises when a blood glucose measurement is taken and the result is displayed in the wrong measuring unit (i.e., mg/dl instead of mmol/l). If it goes unnoticed, the incorrectly displayed blood glucose value may result in inappropriate therapy decisions, including the administration of insulin, and could potentially result in severe consequences. Such possible consequences range from …